The following panel will be displayed after a click on TEACH MEAN VALUES.
The Teach Panel can be used in every EVALUATION and CALCULATION MODE. The explanation herein uses
the EVALUATION MODE=BEST HIT and the CALCULATION MODE=X/Y INT.
The main panel remains active, and data are automatically picked up from the sensor and are displayed.
A click on the CAPTURE button enters a parameter frame in the table.
The COUNTER display field shown how many frames have already been recorded.
The last frames that were entered in the table can be deleted again by pressing UNDO.
RESET TABLE sets the whole table back to 0.
After every activation of CAPTURE, UNDO, or RESET TABLE, the mean values for the individual parameters are
calculated automatically and are displayed in the mean value table.
Furthermore, a maximum color deviation d(X/Y) and a maximum intensity deviation dINT for the average values
will be formed.
When the TEACH TO TABLE button is pressed, the respective mean values are taught to the row in the TEACH
TABLE that is selected under ROW No.: (Æ 2.3 Tab TEACH TABLE).
Teaching of the circular tolerance and of the intensity tolerance can be set by way of TEACH CTO WITH and
TEACH ITO WITH, respectively.
- If the setting is VALUE CTO, the value that is set under VALUE CTO will be taught (ditto intensity).
- If the setting is d(X/Y), the value that is determined under d(X/Y) will be taught (ditto intensity).
- If the setting is d(X/Y) + VALUE CTO, the value that is determined under d(X/Y) plus VALUE CTO will be
taught (ditto intensity).
- With NO CHANGE, the value that is set in the TEACH TABLE will remain unchanged.
A click on the CLOSE TEACH PANEL button will take you back to the main panel.